§ 40.95 Quality assurance plans for non-evidential screening devices.
Wording first on file in the 1996 annual edition, superseded by the 2001 annual edition.
(a) In order to be used for alcohol screening tests subject to this part, a non-evidential screening device shall have an approved quality assurance plan (QAP) developed by the manufacturer and approved by the National Highway Traffic Safety Administration (NHTSA).
(1) The plan shall designate the method or methods to be used to perform quality control checks; the temperatures at which the non-evidential screening device shall be stored and used, as well as other environmental conditions (e.g., altitude, humidity) that may affect the performance of the device; and, where relevant, the shelf life of the device.
(2) The QAP shall prohibit the use of any device that does not pass the specified quality control checks or that has passed its expiration date.
(b) The manufacturers’ instructions on or included in the package for each saliva testing device shall include directions on the proper use of the device, the time frame within which the device must be read and the manner in which the reading is made.
(c) The employer and its agents shall comply with the QAP and manufacturer's instructions for each non-evidential screening device it uses for alcohol screening tests subject to this Part.